MDCG 2019-11: When Is Software a Medical Device?

Whether software is a medical device is decided not by the technology but by the manufacturer’s intended purpose. The MDCG 2019-11 guidance describes how to check this. Get it wrong and you either place a product on the market without the required conformity assessment or plan an approval you never needed.

MDCG 2019-11 in brief

MDCG 2019-11 is the guidance of the EU Medical Device Coordination Group on the qualification and classification of software under the Medical Device Regulation MDR (EU) 2017/745 and the In Vitro Diagnostic Regulation IVDR (EU) 2017/746. The revised version Rev.1 was published in June 2025. The guidance is not legally binding, but authorities and notified bodies use it as their benchmark.

Simplified MDCG 2019-11 path to decide whether software is a medical device, with MDR Rule 11 classes
Simplified qualification path per MDCG 2019-11 Rev.1 and the classes under MDR Rule 11.

Step 1: Is the software a medical device?

The guidance uses a decision tree for qualification. The key questions:

  1. Is it software as defined in the guidance, that is, a set of instructions that processes input data and creates output data?
  2. Does the software act on data beyond storage, archiving, communication or simple search?
  3. Is it for the benefit of individual patients?
  4. Does the intended purpose meet the definition of a medical device in Article 2(1) MDR, such as diagnosis, monitoring, prediction, prognosis, treatment or alleviation of disease?

If the answer to all four is yes, the product is medical device software (MDSW). If the information is derived from in vitro examination of specimens, check whether the IVDR applies instead of the MDR.

Examples of qualification

SoftwareTypical qualificationReason
Appointment scheduling for medical practicesnot a medical deviceno medical purpose for individual patients
Archive that stores image data and displays it unchangedusually not a medical devicestorage and display only
Software that calculates measurements from image data for treatment planningusually MDSWprocesses data for a therapeutic decision
Calculator that suggests a dose from patient dataMDSWsupports a therapeutic decision for an individual patient
Simplified examples. The manufacturer’s specific intended purpose always decides.

Standalone software, part of a device or a module

MDSW can be standalone, for example a cloud application that evaluates image data. Software that drives or influences the use of a physical medical device is assessed together with that device.

If an application consists of several modules, not every module has to meet the MDR requirements. The guidance allows modules with a medical purpose to be separated from other modules. The manufacturer has to define and document the boundaries between the modules clearly.

Step 2: Which risk class applies? Rule 11

For standalone MDSW, the risk class usually follows from Rule 11 in Annex VIII of the MDR:

Purpose of the softwareClass
Provides information used to take diagnostic or therapeutic decisionsIIa
as above, where decisions may cause a serious deterioration of health or a surgical interventionIIb
as above, where decisions may cause death or an irreversible deterioration of healthIII
Monitors physiological processesIIa
Monitors vital physiological parameters where variations could result in immediate danger to the patientIIb
All other softwareI

In practice, a large share of MDSW ends up in Class IIa or higher. From Class IIa, a notified body is involved in the conformity assessment.

What Rev.1 means for existing classifications

Rev.1 updates the 2019 version with revised explanations and examples. If your classification rationale is based on the old version, check it against Rev.1 and record that review in the technical documentation.

What comes after classification

Once the MDR class is clear, development follows the state of the art, usually IEC 62304. Its software safety classes A, B and C are a separate classification and determine which development process and documentation are required. Risk management per ISO 14971, usability per IEC 62366-1 and a quality management system per ISO 13485 complete the picture.

How SimQ supports you

SimQ develops regulatory-compliant MedTech software. The company has been ISO 13485 certified since 2018 and supports medical device manufacturers from the intended purpose through development along IEC 62304 to release. With SimQ OSA, SimQ developed a product combining AI and in-silico technologies and brought it through CE marking as a medical device.

Learn more about Software Development for MedTech. If you plan to work with an external partner, read our guide to outsourcing medical device software development. To discuss your project, book a first call.

Frequently asked questions

Is every health app a medical device?

No. The intended purpose decides. Apps for general wellbeing or pure administration are usually not medical devices.

Who defines the intended purpose?

The manufacturer. It must be reflected in labeling, instructions for use and promotional material. Claims that go beyond the documented intended purpose can change the qualification.

Does MDCG 2019-11 also cover in vitro diagnostics?

Yes. The guidance covers software under both the MDR and the IVDR and helps draw the line between the two regulations.

Is Class I software possible without a notified body?

For Class I software, the manufacturer usually declares conformity itself. The requirements for technical documentation, risk management and the software life cycle still apply.

Note: This article summarizes the guidance in simplified form and does not replace regulatory advice. The MDR and MDCG 2019-11 Rev.1 prevail.